S100B ELISA 试剂盒
-
- 抗原 See all S100B ELISA试剂盒
- S100B (S100 Calcium Binding Protein B (S100B))
- 抗原表位
- Soluble
-
适用
- 大鼠
- 检测方法
- Colorimetric
- 实验类型
- Sandwich ELISA
- 检测范围
- 15.6-1000 pg/mL
- 最低检测浓度
- 15.6 pg/mL
- 应用范围
- ELISA
- 原理
- This immunoassay kit allows for the in vitro quantitative determination of rat S-100B concentrations in serum, plasma and other biological fluids.
- 样品类型
- Plasma, Serum
- Analytical Method
- Quantitative
- 特异性
- This assay recognizes recombinant and natural rat S-100B.
- 交叉反应 (详细)
- No significant cross-reactivity or interference was observed.
- 灵敏度
-
< 3.9 pg/mL
The sensitivity of this assay, or Lower Limit of Detection (LLD) was defined as the lowest detectable concentration that could be differentiated from zero. - 产品特性
- Rattus norvegicus,Rat,Protein S100-B,S-100 protein beta chain,S-100 protein subunit beta,S100 calcium-binding protein B,S100b
- 组件
- Reagent (Quantity): Assay plate (1), Standard (2), Sample Diluent (1×20ml), Assay Diluent A (1×10ml), Assay Diluent B (1×10ml), Detection Reagent A (1×120 μl), Detection Reagent B (1×120 μl), Wash Buffer(25 x concentrate) (1×30ml), Substrate (1×10ml), Stop Solution (1×10ml), Plate sealer for 96 wells (5), Instruction (1)
- 试剂未包括
- Microplate reader. Pipettes and pipette tips. EP tube Deionized or distilled water.
- Featured
- Discover our best selling S100B ELISA Kit
- Top Product
- Discover our top product S100B ELISA Kit
-
-
- 样本量
- 100 μL
- 板类型
- Pre-coated
- 实验流程
- The microtiter plate provided in this kit has been pre-coated with an antibody specific to S-100B. Standards or samples are then added to the appropriate microtiter plate wells with a biotin-conjugated polyclonal antibody preparation specific for S-100B. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. Then a TMB substrate solution is added to each well. Only those wells that contain S-100B, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of a sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450 nm 2 nm. The concentration of S-100B in the samples is then determined by comparing the O.D. of the samples to the standard curve.
- 试剂准备
-
Bring all reagents to room temperature before use. Wash Buffer - If crystals have formed in the concentrate, warm to room temperature and mix gently until the crystals have completely dissolved. Dilute 30 mL of Wash Buffer Concentrate into deionized or distilled water to prepare 750 mL of Wash Buffer. Standard - Reconstitute the Standard with 1.0 ml of Sample Diluent. This reconstitution produces a stock solution of 1,000 pg/mL. Allow the standard to sit for about 10 minutes with gentle agitation prior to making serial dilutions (Making serial dilution in the wells directly is not permitted). The undiluted standard serves as the highest standard (1,000 pg/mL). The Sample Diluent serves as the zero standard (0 pg/mL). pg/mL 1,000 500 250 125 62.5 31.2 15.6 0 Detection Reagent A and B - Dilute to the working concentration using Assay Diluent A or B (1:100), respectively.
- 样品收集
- Serum - Use a serum separator tube and allow samples to clot for 30 minutes before centrifugation for 20 minutes at approximately 1000 g. Remove serum and assay immediately or aliquot and store samples at -20 or -80 . Plasma - Collect plasma using EDTA or heparin as an anticoagulant. Centrifuge samples for 15 minutes at 1000 g at 2 - 8 within 30 minutes of collection. Store samples at -20 or -80 . Avoid repeated freeze-thaw cycles. Other biological fluids - Remove particulates by centrifugation and assay immediately or aliquot and store samples at -20 or -80 . Avoid repeated freeze-thaw cycles. Note: Serum and plasma to be used within 7 days may be stored at 2-8 , otherwise samples must stored at -20 ( 1 month) or -80 ( 2 months) to avoid loss of bioactivity and contamination. Avoid freeze-thaw cycles. When performing the assay slowly bring samples to room temperature.
- 实验流程
-
Allow all reagents to reach room temperature (Please do not dissolve the reagents at 37 directly.). All the reagents should be mixed thoroughly by gently swirling before pipetting. Avoid foaming. Keep appropriate numbers of strips for 1 experiment and remove extra strips from microtiter plate. Removed strips should be resealed and stored at 4 until the kits expiry date. Prepare all reagents, working standards and samples as directed in the previous sections. Please predict the concentration before assaying. If values for these are not within the range of the standard curve, users must determine the optimal sample dilutions for their particular experiments.
1. Add 100 of Standard, Blank, or Sample per well. Cover with the Plate sealer. Incubate for two hours at 37 .
2. Remove the liquid of each well, don ’ t wash.
3. Add 100 μ l of Detection Reagent A working solution to each well. Cover with the Plate sealer. Incubate for 1 hour at 37 . Detection Reagent A working solution may appear cloudy. Warm to room temperature and mix gently until solution appears uniform. 4
4. Aspirate each well and wash, repeating the process three times for a total of three washes. Wash by filling each well with Wash Buffer (approximately 400 μ l) using a squirt bottle, multi-channel pipette, manifold dispenser or autowasher. Complete removal of liquid at each step is essential to good performance. After the last wash, remove any remaining Wash Buffer by aspirating or decanting. Invert the plate and blot it against clean paper towels.
5. Add 100 μ l of Detection Reagent B working solution to each well. Cover with a new Plate sealer. Incubate for one hour at 37 .
6. Repeat the aspiration/wash process for five times as conducted in step
4. 7. Add 90 μ l of Substrate Solution to each well. Cover with a new Plate sealer. Incubate for 15 - 30 minutes at 37 . Protect from light.
8. Add 50 μ l of Stop Solution to each well. If color change does not appear uniform, gently tap the plate to ensure thorough mixing.
9. Determine the optical density of each well at once, using a microplate reader set to 450 nm.
Important Note:
1. Absorbance is a function of the incubation time. Therefore, prior to starting the assay it is recommended that all reagents should be freshly prepared prior to use and all required strip-wells are secured in the microtiter frame. This will ensure equal elapsed time for each pipetting step, without interruption.
2. Please carefully reconstitute Standards or working Detection Reagent A and B according to the instruction, and avoid foaming and mix gently until the crystals have completely dissolved. The reconstituted Standards can be used only once. This assay requires pipetting of small volumes. To minimize imprecision caused by pipetting, ensure that pipettors are calibrated. It is recommended to suck more than 10 μ l for once pipetting.
3. To ensure accurate results, proper adhesion of plate sealers during incubation steps is necessary. Do not allow wells to sit uncovered for extended periods between incubation steps. Once reagents have been added to the well strips, DO NOT let the strips DRY at any time during the assay.
4. For each step in the procedure, total dispensing time for addition of reagents to the assay plate should not exceed 10 minutes.
5. To avoid cross-contamination, change pipette tips between additions of each standard level, between sample additions, and between reagent additions. Also, use separate reservoirs for each reagent.
6. The wash procedure is critical. Insufficient washing will result in poor precision and falsely elevated absorbance readings.
7. Duplication of all standards and specimens, although not required, is recommended.
8. Substrate Solution is easily contaminated. Please protect it from light. - 结果分析
-
Average the duplicate readings for each standard, control, and samples and subtract the average zero standard optical density. Create a standard curve by reducing the data using computer software capable of generating a four parameter logistic (4-PL) curve-fit. As an alternative, construct a standard curve by plotting the mean absorbance for each standard on the x-axis against the concentration on the y-axis and draw a best fit curve through the points on the graph. The data may be linearized by plotting the log of the S-100B concentrations versus the log of the O.D. and the best fit line can be determined by regression analysis. It is recommended to use some related software to do this calculation, such as curve expert 13.0. This procedure will produce an adequate but less precise fit of the data. If samples have been diluted, the concentration read from the standard curve must be multiplied by the dilution factor.
- 限制
- 仅限研究用
-
- 注意事项
-
1. The kit should not be used beyond the expiration date on the kit label.
2. Do not mix or substitute reagents with those from other lots or sources.
3. If samples generate values higher than the highest standard, further dilute the samples and repeat the assay. Any variation in standard diluent, operator, pipetting technique, washing technique,incubation time or temperature, and kit age can cause variation in binding.
4. This assay is designed to eliminate interference by soluble receptors, ligands, binding proteins, and other factors present in biological samples. Until all factors have been 3 tested in the Immunoassay, the possibility of interference cannot be excluded.
5. Limited by the current condition and scientific technology, we can't completely conduct the comprehensive identification and analysis on the raw material provided by suppliers. So there might be some qualitative and technical risks to use the kit. - 储存条件
- 4 °C/-20 °C
- 储存方法
- The Standard, Detection Reagent A, Detection Reagent B and the 96-well strip plate should be stored at -20 °C upon being received. The other reagents can be stored at 4 °C.
-
-
Protective effects of naloxone in two-hit seizure model." in: Epilepsia, Vol. 51, Issue 3, pp. 344-53, (2010) (PubMed).
: "
-
Protective effects of naloxone in two-hit seizure model." in: Epilepsia, Vol. 51, Issue 3, pp. 344-53, (2010) (PubMed).
-
- 抗原 See all S100B ELISA试剂盒
- S100B (S100 Calcium Binding Protein B (S100B))
- 别名
- S100b (S100B 产品)
- 别名
- NEF ELISA Kit, S100 ELISA Kit, S100-B ELISA Kit, S100beta ELISA Kit, S100P ELISA Kit, AI850290 ELISA Kit, Bpb ELISA Kit, si:dkey-110p14.2 ELISA Kit, wu:fq18b10 ELISA Kit, zgc:92776 ELISA Kit, S100 calcium binding protein B ELISA Kit, S100 protein, beta polypeptide, neural ELISA Kit, S100 calcium binding protein, beta (neural) ELISA Kit, S100B ELISA Kit, S100b ELISA Kit, s100b ELISA Kit
- 背景
- S-100B is a member of highly homologous Ca2+ binding proteins family that possess two EF-hand motifs. The family of S100 proteins consists of 19 members. Most S100 proteins exist as dimers (frequently homodimers) within cells. Exclusively expressed in vertebrates, S100 is implicated in various intracellular and extracellular regulatory activities. Studies indicate that S100 proteins are involved in the inhibition of protein phosphorylation, inhibition of cytosceletal constituent assembly, regulation of Ca2+ homeostasis, stimulation of enzyme activities, and interaction with transcription factors. S100B is abundant in the nervous system where it is predominantly expressed in astrocytes, oligodendrocytes and Schwann cells. When secreted by astrocytes, S100B has neurotrophic effects during development and nerve regeneration at physiologic (nanomolar) concentrations. However high (micromolar) concentrations of S100B have shown to be neurotoxic, participating in the physiology of neurodegenerative disorders. The clinical values have been demonstrated in stroke, cerebral complications association with cardiac arrest and in patients with severe as well as minor head injury.
- 途径
- Regulation of Muscle Cell Differentiation, Positive Regulation of Immune Effector Process, Toll-Like Receptors Cascades, Regulation of long-term Neuronal Synaptic Plasticity, S100 Proteins
-